OPERATING MANUALTS - 4Thermal Microscope StageCONTENTSSection Page1.0 GENERAL DESCRIPTION 2
TS-4 THERMAL MICROSCOPE STAGE1.0 GENERAL DESCRIPTIONThe TS-4 Thermal Microscope Stage consists of:1.1 The STAGE, which is available in several config
2.0 SETTING UP THE THERMAL MICROSCOPE STAGESections 2.1 to 2.5 describe setup of the TS-4 with a Pump and Tank Unit, PTU-3. If you areusing running w
2.3 ATTACH AC LINE CORDThis module is on the back panel, bottom left. There are four settings; 100V for unusually lowpower situations, 120V, 220V an
The specimen or slide is placed on the upper surface of the heat transfer plate and held in positionwith the stainless steel clips provided. A sensor
3.0 OPERATING INSTRUCTIONS3.1 Check that SENSOR INPUT is switched to B. The digital display will show the temperatureof the stage.Depress the RUN/SET
4.0 SPECIFICATIONS4.1 CONTROLLEROperating Range: -20oto +60oCControl Accuracy: +0.1oCDigital Readout Resolution: 0.1oCAccuracy: 0.1oC, +
5.0 MAINTENANCE, WARRANTY AND SERVICE5.1 MAINTENANCEThe microscope stage needs no maintenance at all. It may be cleaned as necessary with a softcloth,