Physitemp-instruments TS4 User Manual

Browse online or download User Manual for Sensors Physitemp-instruments TS4. Physitemp Instruments TS4 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 9
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
OPERATING MANUAL
TS - 4
Thermal Microscope Stage
CONTENTS
Section Page
1.0 GENERAL DESCRIPTION 2
2.0 SETTING UP THE THERMAL 3
MICROSCOPE STAGE
3.0 OPERATING INSTRUCTIONS 6
4.0 SPECIFICATIONS 7
5.0 MAINTENANCE, WARRANTY 8
AND SERVICE
EB TS4 1198
Page view 0
1 2 3 4 5 6 7 8 9

Summary of Contents

Page 1 - Thermal Microscope Stage

OPERATING MANUALTS - 4Thermal Microscope StageCONTENTSSection Page1.0 GENERAL DESCRIPTION 2

Page 2 - TS-4 THERMAL MICROSCOPE STAGE

TS-4 THERMAL MICROSCOPE STAGE1.0 GENERAL DESCRIPTIONThe TS-4 Thermal Microscope Stage consists of:1.1 The STAGE, which is available in several config

Page 3

2.0 SETTING UP THE THERMAL MICROSCOPE STAGESections 2.1 to 2.5 describe setup of the TS-4 with a Pump and Tank Unit, PTU-3. If you areusing running w

Page 4

2.3 ATTACH AC LINE CORDThis module is on the back panel, bottom left. There are four settings; 100V for unusually lowpower situations, 120V, 220V an

Page 5

The specimen or slide is placed on the upper surface of the heat transfer plate and held in positionwith the stainless steel clips provided. A sensor

Page 6 - OR IF BUZZER ALARM IS HEARD

3.0 OPERATING INSTRUCTIONS3.1 Check that SENSOR INPUT is switched to B. The digital display will show the temperatureof the stage.Depress the RUN/SET

Page 7

4.0 SPECIFICATIONS4.1 CONTROLLEROperating Range: -20oto +60oCControl Accuracy: +0.1oCDigital Readout Resolution: 0.1oCAccuracy: 0.1oC, +

Page 8

5.0 MAINTENANCE, WARRANTY AND SERVICE5.1 MAINTENANCEThe microscope stage needs no maintenance at all. It may be cleaned as necessary with a softcloth,