OPERATING MANUALTHERMALERT MONITORING THERMOMETERTH-5CONTENTSSection1.0 INTRODUCTION2.0 INITIAL INSTALLATION3.0 P
Here is a useful rough rule: Heat leakage effects are substantially reduced when an amount of probe equal to 10 or more sheath diameters is immersed o
10.0 REPAIRS AND RECALIBRATION10.1 In the event that a Thermalert Monitoring Thermometer is to be returned for repair andrecalibration, please pac
OPERATING INSTRUCTIONSTHERMALERT MONITORING THERMOMETER, TH-51.0 The Physitemp Clinical Monitoring Instruments, Thermalerts TH-5 and TH-8, are compact
4.0 DIRECTIONS FOR USE4.1 Connect any Physitemp type T thermocouple probe to the blue polarized socket on the rear of the instrument.4.2 Switch O
6.0 CHECKING CALIBRATION6.1 The Thermalert TH-5 thermometer incorporates an automatic zeroing amplifier which producesnear-perfect stability of ca
6.6 Checking high end of temperature range:Prepare a water bath at approximately 45°C. Follow measurement instructions given in Section 6.5.6.7 A
7.0 USE OF THERMALERT WITH A CHART RECORDER7.1 Analog output is optional on Model TH-5.WARNING: Use of AC-powered equipment with Physitemp Thermome
8.0 SPECIFICATIONSTemperature Range 25°C - 45°CResolution 0.1°CRepeatability 0.1°CCalibration Conform
9.0 TEMPERATURE MEASUREMENT WITH THERMOCOUPLE SENSORS9.1 The thermocouple is a simple and widely accepted device for measuring temperature. Itcomp