Physitemp-instruments BFS-TC User Manual

Browse online or download User Manual for Sensors Physitemp-instruments BFS-TC. Physitemp Instruments BFS-TC User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 8
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
APPENDIX 1
OPERATING INSTRUCTIONS FOR PHYSITEMP
PUMP AND TANK UNIT, PTU-3
1. Unscrew cap. Fill reservoir with 5 gallons of distilled water.
2. Connect extension tubing to the back of the controller and the connections on the tank. Direction
of the flow is not important. All water fittings are automatically self-sealing when disconnected to pre-
vent water spills.
3. Connect the AC line cord to the receptacle on the pump and to the outlet on the rear of the con-
troller. This ensures that water is always flowing when the controller is on.
4. Switch on the controller.
GENERAL INSTRUCTIONS
The use of distilled water is recommended. This avoids discoloration of the tubing due to organic matter
in untreated water. A purification agent can be added to the water. Any commercially available dehumid-
ifier treatment in liquid, powder or tablet form may be used.
If the stage will be operated for long periods, water temperature may gradually increase. Ice cubes can
be added to the tank to maintain low temperature.
The tank should not be placed more than three feet below the stage itself.
When filling the reservoir for the first time, or when refilling it, a small amount of air may become
trapped in the pump housing and cause intermittent noise, cavitation and a reduced flow rate. If this
occurs, screw the cap firmly onto the tank and tip it at an angle of approximately 45 degrees to the hori-
zontal in all four directions as shown below. This will eliminate air in the pump housing and restore
flow to its maximum.
8
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - GENERAL INSTRUCTIONS

APPENDIX 1OPERATING INSTRUCTIONS FOR PHYSITEMP PUMP AND TANK UNIT, PTU-31. Unscrew cap. Fill reservoir with 5 gallons of distilled water.2. Connect ex

Page 2

7.0 MAINTENANCE, WARRANTY AND SERVICE7.1 MAINTENANCEThe stage needs no maintenace at all. It may be cleaned as necessary with a soft cloth, water or d

Page 3 - 6.0 SPECIFICATIONS

6.0 SPECIFICATIONS6.1 CONTROLLEROperating Range: -40° to +60°C(+100°C in Extended Range versions)Control Accuracy: 0.1°CDigital Readout Resolution: 0.

Page 4 - 4.0 TYPICAL STAGE PERFORMANCE

4.0 TYPICAL STAGE PERFORMANCEThe degree of cold obtained is dependent on the temperature of the cooling water. The stage temperatureshould reach 55-65

Page 5 - 3.7 WARNING LIGHT

3.0 OPERATING INSTRUCTIONS3.1 Check that SENSOR INPUT is switched to B. The digital display will show the temperatureof the stage.3.2 Depress the RUN/

Page 6 - 2.5 TURN ON WATER SUPPLY

2.2 CONNECT THE STAGE TO THE CONTROLLERConnect the two water tubes from the stage to the WATER connection on the front panel of the con-troller. These

Page 7 - BFS/TC SERIES THERMAL STAGES

BFS/TC SERIES THERMAL STAGES1.0 GENERAL DESCRIPTION1.1 Physitemp manufactures three freezing stages for use with microtomes. All stages operate in the

Page 8 - BFS-3TC, BFS-5TC, BFS-30TC

OPERATING MANUALBFS-TC FREEZING STAGESfor microtomesBFS-3TC, BFS-5TC, BFS-30TCCONTENTSSECTION PAGE1.0 GENERAL DESCRIPTION 22.0 SETTING UP THE THERMAL

Comments to this Manuals

No comments